Vaisala HMP260 User Manual
Page 80

HMP260 SERIES
HMP260-U017en-1.5
Appendix 1: Serial commands
FDATE
Adding date to output
FDATE xxx
xxx
=
ON or OFF
When FDATE is activated, the current date is output at the beginning of the output line.
The time is set with command DATE. After RESET or power on the current date is
1991-01-01.
Activating the date output
>fdate on
Form. date : ON
>r
1995-03-10 RH= 21.1 %RH T= 26.0 'C
1995-03-10 RH= 21.3 %RH T= 26.0 'C
1995-03-10 RH= 23.1 %RH T= 26.0 'C
1995-03-10 RH= 22.2 %RH T= 26.0 'C
1995-03-10 RH= 20.6 %RH T= 26.0 'C
...
Inactivating the date output
>fdate off
Form. date : OFF
>r
RH= 20.2 %RH T= 26.0 'C
RH= 19.9 %RH T= 26.0 'C
RH= 19.8 %RH T= 26.0 'C
RH= 19.7 %RH T= 26.0 'C
RH= 19.7 %RH T= 26.0 'C
...
SERI
Serial bus settings
SERI b p d s x
b
=
bauds (300, 600, 1200, 2400, 4800, 9600)
p
=
parity (n = none, e = even, o = odd)
d
=
data bits (7 or 8)
s
=
stop bits (1 or 2)
x
=
duplex (H = half, F = full)
Giving the command on its own outputs the current settings:
SERI
4800 E 7 1 FDX