Press any key and enter the second point value and press .

RH : xx.x Ref2 ? yy.y

The stabilization of the sensor can be monitored by entering c
instead of the reference value.

6.1.3.2

Using display commands

Disconnect the security lock jumper!

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

Place the sensor head into the calibration hole of the LiCl bottle in the
humidity calibrator.

Wait for 10 minutes.

Select Cali in the main menu and then RH cal; select Sensor
changed. Change the first point reading with the arrow keys and
press ENT.

Place the sensor head into the calibration hole of the NaCl bottle in
the humidity calibrator.

Wait for 10 minutes.

If necessary, change the second point reading with the arrow keys and
press ENT.

6.1.3.3

Using LED commands

Disconnect the security lock jumper!

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

Place the sensor head into the calibration hole of the LiCl bottle in the
humidity calibrator.

Wait for 10 minutes.

Connect an ammeter/voltmeter to the analogue outputs (connector
X2). Give command

l¡¡¡. At the first calibration point the LED

on the left flashes; adjust the first point with the arrow switches to the
value given in the calibration table (Chapter 6.1.4) and press ENT
switch.

Place the sensor head into the calibration hole of the NaCl bottle in
the humidity calibrator.

Wait for 10 minutes.