HMP260 SERIES

HMP260-U017en-1.5

Appendix 1: Serial commands

FCRH

Relative humidity calibration after sensor change

Disconnect the security lock jumper!

FCRH

After humidity sensor change the transmitter must be calibrated using this command and
the calibration must be done at two reference points. The calibration is performed as
follows:

FCRH

RH : 1.90

Ref1 ?

11.3

Press any key when ready ...

RH : 76.30

Ref2 ?

74.9

The command can also be divided into two commands, so the computer can be used
for other purposes while waiting for the sensor to stabilize to the higher humidity.

FCRH 1

RH : 1.90

Ref1 ?

11.3

FCRH 2 ...

RH : 76.30

Ref2 ?

74.9

If the stabilization of the sensor to the humidity in the calibrator needs to be monitored,
the measurement output can be repeated by giving command c at Ref1 and Ref2:

FCRH

RH : 12.00

Ref1 ? c

RH : 11.70

Ref1 ? c

RH : 11.50

Ref1 ? 11.3

Press any key when ready ...

RH : 76.00

Ref2 ? 75.5

CT

Temperature calibration

Disconnect the security lock jumper!

CT

With this command the transmitters can be calibrated against an accurate reference,
such as a Pt 100 simulator. A two-point calibration is performed as follows:

CT

T : 0.80 Ref1 ?

0.0

Press any key when ready ...

T : 56.20

Ref2 ?

55.0