29
5.3.6.1
Setting the serial interface operation mode
•
Select Mode in the main menu; the following is displayed:
•
Select Serial output:
•
The currently valid setting flashes. Select the desired mode with the
arrow keys and press ENT. After this the programme returns to the
Mode Menu.
•
When Run mode is selected, the currently valid output interval is dis-
played:
The output interval setting can be changed as follows:
•
press CL
•
the number starts flashing
•
if the interval needs to be changed, press CL again and enter the new
interval; otherwise press ENT
•
the unit (seconds or hours) starts flashing
•
the unit can be changed with the arrow keys and acknowledged with
ENT
•
after this the programme returns to Mode menu
5.3.7
Others
5.3.7.1
Setting the measurement integration time
By lengthening the measurement integration time any stray changes in the out-
put can be filtered out: the transmitter calculates the average of a number of
measurement cycles defined by the user.