3. Insert the body of the sensor head into the fitting body and screw the

nut manually to the fitting body until the connection feels tight.

4. Mark both the fitting body and the nut hex.

A pen

fitting body

nut

5. Tighten the nut a further 30° (1/12 turn) or if you have a torque span-

ner tighten it with a torque of 80 ±10 Nm.

NOTE

After detachment the nut must be tightened without
increased effort.

30°

6. The tightening cone of the fitting body has to be cleaned and greased

after each tenth detachment. Use high-vacuum grease (Down Corning,
Europe) or a similar grease.

clean cotton stick

tightening cone

sealing
washer

fitting body

The sealing washer has to be changed every time the fitting body is
detached.