Vaisala HMP260 User Manual
Page 86

HMP260 SERIES
HMP260-U017en-1.5
Appendix 1: Serial commands
or pressure is set to zero using XPRES. After this the pressure stored using command
PRES is valid again.
CDATE
Entering calibration date
CDATE xxxxxx
xxxxxx =
calibration date (000101...991231)
When the latest calibration date has to kept in memory, it is entered as follows:
CDATE 940506
If the command is given without the date, the transmitter outputs the latest calibration
already in memory.
CDATE
940420
The date can be given in any format; however, the maximum number of numbers is six.
DATE
Setting the date
DATE
For example, to enter a new date:
DATE
Current date is 1993-01-30
Enter new date (yyyy-mm-dd) : 1993-06-12
When the current date is asked, the new date is passed with
TIME
Setting the time
TIME
For example, to enter a new time:
TIME
Current time is 01:35:54
Enter new time (hh:mm:ss) : 13:25:56
When the current time is asked, the new time is passed with