7

NOTE

When using the support bar the lead-through piece and
the space between the duct wall and the flange must be
sealed e.g. with silicon. The sealings must fulfil EEx
requirements.

The sensor head can also be installed vertically.

when a bushing is used, its size is
selected according to the diameter
of the sensor head; the diameter

of the cable increased by using
e.g. tape at the bushing

to be sealed

(loop type)

clamp

mounting the
sensor head
on a support

HAZARDOUS
AREA

SAFE AREA

Fig. 3.3

Vertical installation of the HMP263 sensor head

3.2.2.

HMP264 transmitter

The HMP264 is supplied with a nut, a fitting body and a sealing washer.
During handling the fitting body and the nut should remain in place on the
body of the sensor head to prevent damage to the highly polished surface.

To achieve a leak-tight assembly:

1. Remove the fitting body from the nut and the sensor head.

2. Fasten the fitting body to the chamber wall. Tighten the fitting body

into the threaded sleeve with a torque spanner. The tightening torque
is 150 ±10 Nm.

tightening cone

sealing washer

fitting body

nut

SAFE AREA

HAZARDOUS AREA