Vaisala HMP260 User Manual
Page 87

HMP260 SERIES
Appendix 1: Serial commands
HMP260-U017en-1.5
VERS
Name and version of the programme
VERS
For example:
VERS
HMPS 260 / x.yy
where x.yy is the programme version.
?
Outputting the transmitter settings
?
For example:
?
HMPS 260 / 1.01
CPU serial nr
: 0
Keyboard type
: 0
Address
: 7
Output units
: metric
Baud P D S : 4800 E 7 1 FDX
Serial mode
: STOP
Output intrv.
: 0 min
Mtim
: 32
Pressure
: 1013.25
Analog outputs
Ch1 0.00 ... 10.00 V
Ch2 0.00 ... 10.00 V
Ch1 ( RH )
lo
0.000 %RH
Ch1 ( RH )
hi
100.000 %RH
Ch2 ( T )
lo
-20.000 'C
Ch2 ( T )
hi
180.000 'C
Transducer :
PRB serial nr
: 0
Calibr. date
: 0
??
Outputting the transmitter settings also in POLL mode
??
Command ?? outputs the same information as command ? but it works also when the
transmitter has been set to POLL mode. However, if there are more than one
addressed transmitters connected to the serial bus, they all will respond at the same time
and the output on the screen will be chaotic.