Instruction Manual

CyberComm 6000

26

3.3 Select existing standardization data of CyberComm for

measurement


Follow the steps below, if you wish to use standardization data, available in the
CyberComm. This includes standardization done by you previously for the same
meter and for the same channel/parameter or standardization done by other users for
the same meter and same channel/parameter.
To begin:
1. Make sure Multi-channel check box is not selected in the Measure control tab.
2. Select the channel and its parameter that you intend to standardize.
3. Click Setup tab and set appropriate standardization options.

ƒ

For pH standardization, click on Setup tab and select the Buffer Group.

ƒ

For conductivity standardization, click on Setup tab and select appropriate

Standard Recognition.

ƒ

For DO standardization, click on Setup tab and select appropriate

Standardization mode.

4. Click on Standardization tab in the workspace.
5. Electrode Configuration Dialog appears. (Figure 24)
6. in User drop-down box, select :

ƒ

You user ID: if you have already performed standardization with this

project or any other project and you wish to use that info

ƒ

Any other user ID: if any other user has already performed

standardization and you wish to use that info.

7. The Project Name (Job Name) drop-down box lists all the available projects for

the selected user. Select required project from the list

8. The Electrode Serial No drop-down box lists available electrode list for the

selected project.

9. Select required electrode’. Existing Standardization grid shows standardization

data for the selected channel, selected parameter, for the selected
standardization buffer/range and for the selected electrode. Depending on the
selected channel/parameter, it also shows related information such as
standardization date, slope, offset, buffer expiry date etc.

10.

If you wish to use the standardization data shown in the grid for your
measurement, click Load Selected Standardization button.

11. The Electrode Configuration Dialog exists. If you have selected standardization

data from another project, CyberComm copies the selected electrode to the
current project. The selected standardization data is loaded to the selected
channel/parameter and the CyberComm is ready for measurement using the
selected standardization.

Note: If the selected electrode of the project does not have a matching
standardization data for the selected channel/parameter/buffer/range, the
Existing Standardization grid does not show any data.