51

ERROR

MESSAGE

(BLINKING)

INDICATES

PROBABLE CAUSE

CORRECTIVE ACTION

Incorrect

readings

1. Incorrect

calibration.

2. Wrong

temperature

coefficient %.

3.

Wrong TDS factor.

4. Dirty

electrode.

1. Reset

calibration

and re-calibrate

with correct

parameters.

2. Reset

Temperature

compensation %.

3.

Reset TDS factor.

4.

Clean electrode in

isopropanol. If

electrode is coated

with contaminants,

clean using soft

tissue and re-

calibrate.

Not

responding to

key press

1. HOLD

mode

selected.

2.

Internal program error.

1. Cancel

HOLD

mode.

2. Re-insert

batteries.

Err 1

Memory write error

Return to distributor for

repair or replacement.

Err 2

Memory read error

Return to distributor for

repair or replacement.

Err 3

A/D error

Return to distributor for

repair or replacement.

Err 4

Keypad error

Return to distributor for

repair or replacement.

Err 5

EEPROM read

error

Return to distributor for

repair or replacement.