8.2

Program 1: Common functions

The following 3 options are commonly used:

8.2.1

P1.0: Memory Clear

16 sets of readings can be stored in the meter. Both

conductivity and temperature or TDS and temperature

readings are stored.

ON

Clears all values stored in memory

OFF

Memory values remain in storage

Activation of this option by selecting ON and confirmation of

memory clears all memory values. This is done by toggling

the annunciator to ON using the MI/

!

or MR/

"

key, and

pressing ENTER/RANGE to confirm the selection.

8.2.2

P1.1: %Temperature coefficient
adjustment

The amount of change in conductivity per degree centigrade

is referred to as the Temperature Coefficient, expressed in

%/

°

C at a particular temperature. The meter provides the

options of selecting the appropriate temperature coefficient,

from 0.0 to 10.0 %/

°

C, depending on the type of solution

measured.

The primary display indicates the temperature coefficient in

%/

°

C. You can use the MI/

!

or MR/

"

key to scroll to a

required value from 0.0 to 10.0 %/

°

C. Once you have made

your selection, press ENTER/RANGE key to confirm the

value.

The program will automatically scroll to the next SETUP

parameter. Under default condition, the temperature

coefficient is set to 2.0%/

°

C.

Figure 23: Memory clear display

Figure 24: Temperature

coefficient adjustment setup