31
8.3.3
P2.2: Calibration Reset
P2.2 provides the option to reset all user selectable
functions to the default settings. If you toggle the
annunciator to ON using the MI/
!
or MR/
"
key, and press
ENTER/RANGE key, all data will be reset. RESET clears
all calibration data in memory (of both conductivity
and TDS). You need to power ON the instrument before
proceeding with any other functions.
Under default condition, RESET option is OFF.
NOTE: Data cannot be selectively deleted.
ON
Reset all calibration points
OFF
Reset off
8.3.4
P2.3: Temperature Normalization
The ATC (Automatic Temperature Compensation)
function automatically compensates for the variation
caused by temperature difference during measurement
and at the standard temperature (20 or 25
°
C) and gives
the corrected read-out of the equivalent conductivity of
the solution. Hence, the conductivity/TDS can be
normalized to display the equivalent conductivity/TDS at
either 20 or 25
°
C.
Normalization temperature can be selected between 20 &
25
°
C using the MI/
!
or MR/
"
key. The primary display
shows normalization temperature. Pressing
ENTER/RANGE key registers the selected temperature at
which the conductivity/TDS value is normalized. Default is
at 25
°
C.
Figure 28: The RESET option
Figure 29: The Temperature
Normalization option