TDS calibration: Method 2 (for CyberScan CON 200 only)

1. Select the TDS standard solution with a value

that is closest to the measurements expected in

the sample (in this manual, the KCI standard

solution 1413

µ

S is about 706.5 ppm at 25

°

C).

2.

Pour about 3 cm of the selected sample into two

separate clean containers.

3. Rinse the end of the probe into one of the

containers by gently stirring it in the standard

solution. This helps remove contaminants that

offset the calibration and introduce error.

4. Dip the rinsed probe into the other container

containing the unused buffer solution.

5.

If the Eutech standard solution is used, the correct

Conductivity-to-TDS conversion factors are given

for KCI, NaCI & 442 natural water formulation at

25

°

C under the “Multiply By:” on the label. If

another brand of TDS calibration is used, refer to

the formula:

Factor =

TDS

Conductivity

@ 25

°

C

Select the correct factor according to the protocol

of the method, or that is closest to the solution to

be measured.

6.

Enter into the SETUP mode by pressing the SET

key (note that this mode can be entered only from

the MEAS mode).

Press the ENTER/RANGE key repeatedly until Program 3.0 (TDS factor) is shown.

Figure 15: TDS calibration

procedure Method 2

Press ENTER key
repeatedly until you
come to P3.0.