18

7.

Using the MI/

!

or MR/

"

key, adjust the value of the primary display until it is the same

or close to the selected/calculated value.

Confirm the selected value by pressing the ENTER/RANGE key.

8.

Exit the SETUP mode and return to the MEAS mode by pressing the CAL/MEAS key

upon completion of step 7. Press the MODE key until the TDS mode is entered. The TDS

value will be the buffer solution value for the type of TDS solution selected. The value will

also be temperature compensated and the conductivity-to-TDS conversion input will

ensure greater accuracy over a broader range of TDS curves used.