8.4

Program 3: TDS Setup (in CyberScan CON 200 only)

8.4.1

P3.0 TDS factor

The TDS factor for a particular standard solution is a

multiplication factor which relates the measurement of

conductivity in

µ

S/cm to its equivalent reading in

ppm/mg/l. This factor is unique for each solution.

This option allows the selection of the TDS factor for the

solution to be measured. The TDS factor is shown on the

primary display. Toggle the MI/

!

or MR/

"

key to select

the exact TDS factor. The range of values is from 0.4 to

1.0. Any values less than 0.4 or more than 1.0 would

cause an error. Press the ENTER/RANGE key to confirm

the TDS factor selected. Under default conditions, the

TDS factor is set to 0.5.

Figure 31: The TDS setup