10 TROUBLESHOOTING & ELECTRODE CARE

10.1 Troubleshooting

ERROR

MESSAGE

(BLINKING)

INDICATES

PROBABLE CAUSE

CORRECTIVE ACTION

ERR

Invalid key

sequence

Wrong key press.

Select correct key

ERR

Upper/lower limit

reached

INC/DEC keys presses

beyond the allowable limits.

Reselect parameter

within limits

CAL + ERR

Unable to accept

calibration

1. Incorrect

solution

conductivity/TDS value

entered.

2.

Incorrect TDS factor

entered.

1.

Enter the correct

TDS or conductivity

value (note the

temperature of the

calibration

solution).

2.

Enter the correct

TDS factor.

PRINTER +

ERR

Printing error

1.

Printer is not ready or

unable to receive data.

2.

Printer is improperly

connected.

1.

Ensure that printer

is set to receive

information. Check

also the SETUP

data

communication

settings.

2.

Check the RS

232C port cable

connection.

Unable to

switch on

Battery problem

1.

Batteries not in place.

2. Incorrect

battery

orientation.

3. Weak

batteries.

1. Insert

batteries.

2.

Place the batteries

at the correct

orientation.

3. Replace

with

fresh

batteries or use

A.C. adapter.