Manual Temperature Compensation

IMPORTANT: For manual compensation, you must disconnect the temperature probe. The

ATC annunciator will disappear from the LCD.

1.

Press the MODE key to select TEMP mode.

2. Press the CAL/MEAS key to enter

Temperature calibration mode. The CAL

indicator will appear above the primary

display.

3. The primary display shows the current

conductivity/TDS reading and the

secondary display shows the default

temperature value 25

°

C.

4. Check the temperature of your sample

using an accurate thermometer.

5. Press

the

!

or

"

, or MI/

!

or MR/

"

key to

set the temperature to the measured value

obtained in step 4. To confirm, press

ENTER/RANGE key.

6.

Press the CAL/MEAS key to exit and return

to conductivity measurement mode.

NOTE: To exit this program without confirming

the manual temperature compensation value, DO

NOT press ENTER/RANGE in step 5. Press

CAL/MEAS instead.

Figure 17: Setting the Temperature

Compensation manually