beautypg.com

Printing errors, Sending data to the computer, Printing measurement data – Thermo Fisher Scientific CyberScan CON 10/CON 100/CON 200 User Manual

Page 28: Printing data from memory

background image

25

7.2 Printing

errors

The meter displays an error by blinking the printer

and error annunciators if the printer is not ready to

receive data or if the printer is off. As soon as the

printer is ready, the error display will automatically go

off. While the meter is displaying the printer error, the

user may press the CAL/MEAS key to go back to the

measurement mode.

7.3

Sending data to the computer

To send data to the computer, connect the RS 232C

cable from the base of the meter to the COMM port 1

of the computer. Load and run the CYDATA data

acquisition program ensuring that the parameters of

the settings in the meter and the CYDATA are

identical.

To send data to the computer, press the

key.

7.4 Printing

measurement

data

To print any data that is currently being measured, press the

key. Note that the printing

capability is available for the COND, TDS and TEMP modes.

The data will be printed onto the printer paper or the screen of the CYDATA program

automatically as illustrated in Page 26.

7.5

Printing data from memory

The CyberScan CON 200 meter can print data that is stored in the memory. Firstly, ensure

that the cables are properly connected to either the printer or the computer and the units are

configured to receive the data from the meter.

Change the mode to the mode of measurement from

which data needs to be printed.

Scroll to the data that needs to be printed by pressing

the MR/

"

key. To print, press the

key. The data

displayed will then be sent to the printer or the computer.

An example of the printout is shown in Page 26.

Figure 20: Display of printing

error

Figure 21: Sending data to the

computer

Figure 22: Printing data from

memory