beautypg.com

7 troubleshooting – SHIMPO DT-365E User Manual

Page 26

background image

26

7 Troubleshooting

Symptoms

Factors

Causes

Solution

When the power is turned ON,

"LLLLLL" is indicated, and no

emission occurs.

The battery voltage is

low.

The battery is not

charged.

Charge the battery.

Though the battery is fully

charged, "LLLLLL" is soon

indicated on the display.

The battery capacity is

low.

Short battery life

Battery malfunction

Charging circuit malfunction

Replace the battery.

If the problem has not been

solved after replacing the

battery, ask for repair.

The charging lamp does not

light up when the AC adapter is

connected.

The power does not

reach the main unit.

The AC adapter is not

connected.

Check the connection.

AC adapter malfunction

Ask for repair.

Charging cannot be

completed, and the charging

lamp remains ON.

Charging failed.

Short battery life

Battery malfunction

Charging circuit malfunction

Replace the battery.

If the problem has not been

solved after replacing the

battery, ask for repair.

Emission occurs inconsistently.

Emission sometimes stops.

Internal circuit

malfunction

Internal circuit

malfunction

Ask for repair.

Display or emission does not

change by turning the dial.

Internal circuit

malfunction

Internal circuit

malfunction

Ask for repair.

Information appears on the

digital display, but no emission

occurs.

Internal circuit

malfunction

Internal circuit

malfunction

Ask for repair.

Nothing is indicated on the

display, and no emission

occurs.

Internal circuit

malfunction

Internal circuit

malfunction

Ask for repair.