Martel Electronics BetaGauge 3II User Manual
Page 91
Application Notes
6-6
SYS Mode
There are two methods for configuring an electrical to
pressure calibration test using SYS mode. The
following table indicates which method to use.
Use this method if…
Method 1
Your field instrument has a pressure
output range within the ability of the
pressure module on the BetaGauge II.
Method 2
Your field instrument has a pressure
output range that exceeds the ability of
the pressure module on the BetaGauge
II.
NOTE: While in SYS mode, the BetaGauge II cannot
measure the electrical input signals to the field
instrument. You must manually enter the value
of these signals on the BetaGauge II.
Method 1
Software Instrument Setup
Electrical input (requires manual entry on the
BetaGauge II)
Pressure output
Other Necessary Equipment
Electrical source
- 3001 (94 pages)
- BetaGauge 301 (40 pages)
- BetaGauge 311A-Ex (32 pages)
- BetaGauge 311A (48 pages)
- BetaGauge 330 (48 pages)
- BetaGauge PI (20 pages)
- BetaGauge PIR (24 pages)
- BetaLOG-PI (28 pages)
- BetaLOG-TI (16 pages)
- BetaProbe TI/TI+ (9 pages)
- DPC-30 (12 pages)
- DPC300A (48 pages)
- DMC1410 (108 pages)
- IVC-222HPII (12 pages)
- LC-110/H (50 pages)
- BETA LDT-500 (2 pages)
- M2000 (28 pages)
- MC1010 (46 pages)
- MC1210 (56 pages)
- TC100 (16 pages)
- T150 (8 pages)
- T140 Series (12 pages)
- PTC8010 (36 pages)
- PSC4010 (36 pages)
- MS-420 (1 page)
- MECP500 (2 pages)
- MECP2000 (2 pages)
- MECP10K (4 pages)
- MECP100 (2 pages)
- T130 (2 pages)
- T120 (2 pages)
- MC1200 (50 pages)
- MC1000 (28 pages)
- MasterCAL (140 pages)
- LC100 (20 pages)
- DMC1400 (106 pages)
- 320 (60 pages)
- 230 (106 pages)
- 130 (85 pages)
- QM100V (1 page)
- QM100V-40 (1 page)
- QM110T (2 pages)
- QM120ET (2 pages)
- QM130M (2 pages)