Martel Electronics BetaGauge 3II User Manual
Page 144
![background image](/manuals/369778/144/background.png)
7-3
Doc. Cal. Mode
Connecting Necessary Equipment
Make sure that you have connected all of the necessary
equipment required for the calibration test you want to
perform. This equipment includes the field instrument
you want to calibrate, the BetaGauge II, as well as, any
other equipment necessary to performing a particular
calibration test. For information on how to configure
the equipment, refer to Section 6, Application Notes.
SYS Mode Levels
The SYS mode has three primary levels of operation:
Main
Back Screen
Execute
Main Level
When you first access the BetaGauge II SYS mode, you
are in the Main level. You can use the Main level to
perform the following functions:
Select a tag for testing or calibrating purposes.
View the overall calibration status of a selected tag
and its allowed tolerances.
View specific test point details for a selected tag.
View the results of the last test performed on a
selected tag.
Access the Back Screen level.