Specifications
B-4
Accuracy
Pressure
Range
Accuracy
Overpressure
Notes
Gauge (PSIG)
0 to 0.3 (0 to 20mBar)
±0.1%
400 %
0 to 1 (0 to 67mBar)
±0.05%
400 %
0 to 5 (0 to 350 mBar)
±0.025%
400 %
9
0 to 7.2 (0 to 500 mBar)
±0.035%
300 %
4 & 7
0 to 10 (0 to 700 mBar)
±0.025%
300 %
9
0 to 30 (0 to 2 Bar)
±0.025%
300 %
0 to 50 (0 to 3.5 Bar)
±0.03%
300 %
0 to 100 (0 to 7 Bar)
±0.025%
300 %
0 to 150 (0 to 10 Bar)
±0.035%
200 %
4
0 to 300 (0 to 20 Bar)
±0.025%
200 %
0 to 500 (0 to 34 Bar)
±0.05%
200 %
0 to 1000 (0 to 70 Bar)
±0.05%
200 %
6
0 to 1500 (0 to 100 Bar)
±0.05%
200 %
4
0 to 3000 (0 to 200 Bar)
±0.05%
200 %
0 to 5000 (0 to 340 Bar)
±0.05%
200 %
0 to 10000 (0 to 700 Bar) ±0.05%
120 %
Vacuum (PSIG)
0 to -5 (0 to -350 mBar)
±0.025%
400 %
9
0 to -15 (0 to -1 mBar)
±0.025%
300 %
7
Absolute (PSIA)
0 to 15 (0 to 1 Bar)
±0.025%
300 %
7
0 to 30 (0 to 2 Bar)
±0.025%
300 %
0 to 50 (0 to 3.5 Bar)
±0.03%
300 %
0 to 100 (0 to 7 Bar)
±0.025%
300 %
0 to 300 (0 to 20 Bar)
±0.025%
200 %
Compound (PSIG)
-15 to 15 (-1 to 1 Bar)
±0.025%
300 %
7
-15 to 30 (-1 to 2 Bar)
±0.025%
300 %
Differential (PSID) (Note1)
0 to 5 (0 to 350 mBar)
±0.025%
400 %
5 & 8
0 to 30 (0 to 2 Bar)
±0.025%
300 %
5
0 to 50 (0 to 3.5 Bar)
±0.03%
300 %
5