Martel Electronics BetaGauge 3II User Manual
Page 163
Doc. Cal. Mode
7-22
Accessing the Post-Test Status Screen
After you have calibrated the last test point for the
current tag, the BetaGauge II displays the Post-Test
Status screen. A sample is shown below:
Post-Test Status Screen Description
The Post-Test Status screen is exactly like the Tag
Status screen except for the menu items that appear at
the bottom of the screen.
NOTE: Refer to the Tag Status screen description
(earlier in this section) for a complete
description of the screen data.
From the Post Test Status screen, you can perform the
following functions:
Use the menu items at the bottom of the screen to
indicate whether the field instrument under test was
adjusted (Yes), left alone (No), or replaced (Rep).
Access another system prompt that allows you to
indicate the identification number of the technician
who performed the calibration. This entry is
optional.
Tag 2:
BG2 TAGS Tag 2
Inst. OUT: mA
Inst. IN:
PSI
Status:
PASS
Time:
11:23:59
Date:
10/07/96
Error Tol:
0.0500
Alert Tol:
999999
Next: As Left 1
Adj. Instr.
Yes
No
Rep