Martel Electronics BetaGauge 3II User Manual
Page 153
Doc. Cal. Mode
7-12
Erasing Data for a Specific Tag
To erase all recorded data for a specific tag on the Back
screen, do the following:
NOTE: Recorded data exists only if you have
performed calibration tests for the selected tag.
For information on how to perform a
calibration test, see the procedure titled Tag
Calibration later in this section.
1.
Specify the tag number for the data you want to
erase.
2.
Press the function key corresponding to the menu
item Reset and the bottom of the back screen
appears as shown below:
NOTE: The Reset menu item is displayed only if
calibration reset capabilities are allowed in your
instrument maintenance.
3.
Do one of the following:
To erase all recorded data for the specified tag
and return to the Tag Status screen, press the
function button corresponding to the Yes menu
item.
To cancel the operation and return to the Tag
Status screen, press the function button
corresponding to the No menu item.
NOTE: You can exit the SYS mode by pressing
the CLR key.
Reset Cal.
Yes
No