Martel Electronics BetaGauge 3II User Manual
Page 150
7-9
Doc. Cal. Mode
Viewing the Calibration Results for a Specific
Tag
When you have completed calibrating a specific tag,
you can view the calibration results on the BetaGauge
II.
NOTE: The BetaGauge II will display the calibration
results from the last calibration performed on
the current tag. For information on how to
calibrate test points, see the procedure titled
Tag Calibration later in this section.
From the Tag Status screen use the following procedure
to view the last calibration results for a specific tag:
1.
Specify the tag number for the calibration results
you want to view.
2.
Press the function key corresponding to the menu
item Rslt to display the Review Test Results screen.
A sample is shown below:
3.
If there are additional test points you want to view,
press the function key corresponding to the PgDn
menu item to display the additional test points.
Tag 2:
BG2 TAGS Tag 2
Inst. OUT: mA
Inst. IN:
PSI
Pt.
IN
OUT
Un 1
3.0000
4.0000
Un 2
6.0000
8.0000
Un 3
9.0000
12.000
Un 4
12.000
16.000
Un 5
15.000
20.000
Un 6
12.000
16.000
Review Test Results
PgDn