7-19

Doc. Cal. Mode

Execute Screen Description

The Execute screen is divided into six sections. These
sections are described below in the order that they
appear on the screen from top to bottom:

Tag Number/Area. The current tag number and area is
listed in the top portion of the screen.

NOTE: When using an instrument maintenance

software package, all tags must be assigned to
an area. Refer to your instrument maintenance
software manual for more information.

Input/Output Types. This section lists the field
instrument output type (mA in the example) and the
field instrument input type (PSI in the example).

Calibration Status. The calibration status appears to
the left of each listed test point:

Un. Uncalibrated
C. Calibrated
CE. Calibration Error

Test Point Data. This section shows the expected input
and output measurements for the first three test points
of the current tag.

Actual Reading. This section shows the test point that
is currently under test, as well as, live displays of the
module inputs.

Instrument Input. The bottom section prompts you to
apply the required field instrument input value for the
current test point.