beautypg.com

Operation – KROHNE PTIFLEX 2200 FOUNDATION FIELDBUS User Manual

Page 29

background image

OPERATION

4

29

OPTIFLEX 2200 C

www.krohne.com

09/2013 - 4002573201 - AD FF OPTIFLEX 2200 R01 en

SUB_CODE

Subcode

R

An enumeration specifying the cause of the alert to be reported.

Other

VALUE

Value

R

The value of the associated parameter at the time the alert was

detected.

FD_OFFSPEC_MAP

FD_OFFSPEC_MAP

FD_OFFSPEC_MAP

FD_OFFSPEC_MAP

Offspec Map

Offspec Map

Offspec Map

Offspec Map

RW

This parameter enables or disables conditions to be detected as

active for this alarm category.

Nothing

mapped

Setting:

(00) Check Function

(01) Cond01

(02) Cond02

(03) Cond03

(04) Cond04

(05) Cond05

(06) Snapshot invalid

(07) Audio Signal Offset outside Range

(08) Reference Position outside Range

(09) Pulse Flange Lost

(10) No Probe detected

(11) Reference Lost Error

(12) Oscillator Frequency failure

(13) Temperature Above +75°C

(14) Temperature Below -35°C

(15) Overfill Warning

(16) Level Lost warning

(17) Temperature out of range (warning)

(18) Local operation

(19) Overfill Error

(20) Level lost error

(21) Temperature out of range Error

(22) Converter Voltage Error

(23) Converter ROM Error

(24) Converter RAM Error

(25) Converter EEPROM Error

(26) Sensor not compatible

(27) Internal Communication Error

(28) Sensor Voltage Error

(29) Sensor ROM Error

(30) Sensor RAM Error

(31) Sensor EEPROM Error

Parameter

Parameter

Parameter

Parameter
DD name

Access Description and settings

Initial Value