Operation – KROHNE PTIFLEX 2200 FOUNDATION FIELDBUS User Manual
Page 24
4
OPERATION
24
OPTIFLEX 2200 C
www.krohne.com
09/2013 - 4002573201 - AD FF OPTIFLEX 2200 R01 en
SUB_CODE
Subcode
R
An enumeration specifying the cause of the alert to be reported.
Other
VALUE
Value
R
The value of the associated parameter at the time the alert was
detected.
Setting:
State 0...16
FD_FAIL_MAP
FD_FAIL_MAP
FD_FAIL_MAP
FD_FAIL_MAP
Fail Map
Fail Map
Fail Map
Fail Map
RW
This parameter enables or disables conditions to be detected as
active for this alarm category.
(00)-(31)
Setting:
•
(00) Check Function
•
(01) Cond01
•
(02) Cond02
•
(03) Cond03
•
(04) Cond04
•
(05) Cond05
•
(06) Snapshot invalid
•
(07) Audio Signal Offset outside Range
•
(08) Reference Position outside Range
•
(09) Pulse Flange Lost
•
(10) No Probe detected
•
(11) Reference Lost Error
•
(12) Oscillator Frequency failure
•
(13) Temperature Above +75°C
•
(14) Temperature Below -35°C
•
(15) Overfill Warning
•
(16) Level Lost warning
•
(17) Temperature out of range (warning)
•
(18) Local operation
•
(19) Overfill Error
•
(20) Level lost error
•
(21) Temperature out of range Error
•
(22) Converter Voltage Error
•
(23) Converter ROM Error
•
(24) Converter RAM Error
•
(25) Converter EEPROM Error
•
(26) Sensor not compatible
•
(27) Internal Communication Error
•
(28) Sensor Voltage Error
•
(29) Sensor ROM Error
•
(30) Sensor RAM Error
•
(31) Sensor EEPROM Error
Parameter
Parameter
Parameter
Parameter
DD name
Access Description and settings
Initial Value