beautypg.com

Service, 1 electronics exchange, 1 exchange of the electronics unit – KROHNE UFM 530 HT EN User Manual

Page 54

background image

7

SERVICE

54

UFM 530 HT

www.krohne.com

12/2013 - 4002393201- HB UFM 530 HT R02 en

7.1 Electronics exchange

7.1.1 Exchange of the electronics unit

DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!

WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.

Figure 7-1: Switching off power

1 Remove the back cover using the wrench.
2 Disconnect all cables.

Figure 7-2: Removing display print

1 Remove the front cover with the wrench.
2 Loosen the screws of the display.