6.4.5 Direction of current output (menu 3.04.02)

6.4.6 User-defined unit for volume flow rate and totalizer (menu 3.07.05...3.07.07)

Any user-defined unit can be programmed. This unit is to be defined as a volume
unit per time unit.

• Function 3.07.05: enter the text you want to show at the display.

• Function 3.07.06: enter the amount of the wanted volume units that will fit in a m

3

.

• Function 3.07.07: enter the amount of seconds to equal the wanted time unit.

Example

To program barrels per day

1. Enter "bar/day" in menu 3.07.05.

2. 1 barrel equals 0.159 m

3

, so 1/0.159 = 6.289 barrels are needed to fill 1 m

3

. Enter "6.289" in

menu 3.07.06.

3. 1 day equals 86400 seconds (= 24 x 60 x 60) so enter 8.640E4 in menu 3.07.07 (8.640E4 = 8.640

x 10

4

).

INFORMATION!
This function is only available when ACT FLOW or CORR FLOW is selected in function 3.04.01.
When VOS or GAIN is set, only the forward characteristic applies.

Figure 6-1: Test

1 I: Axis for current output.
2 I

max

(≤ 22 mA): Current output maximum.

3 I

100%

: Current output at 100% scale.

4 Q: Axis for volume flow rate in %.
5 Selected option: BOTH, reverse flow.
6 Selected option: FORWARD.
7 Selected option: F/R SPEC.