Ex NAMUR version

Figure 4-8: Terminals for Ex instrument with NAMUR

Terminal

Function

Specification

1

Neutral mains power supply

100...240 VAC, 24 VAC or 24 VDC

2

Live mains power supply

100...240 VAC, 24 VAC or 24 VDC

3

Combined current output (I) and digital
input (C). Current output (I) includes
HART

®

-Communication.

Current output (I): I ≤ 22 mA,
R

load

≤ 680 Ω, U

max

= 15 VDC.

Digital input (C): low = 0...5 VDC,
high = 15...32 VDC (will be switched off
when current output activated).

4

Pulse / frequency output

I

max

= 150 mA, U

max

= 32 VDC / 24 VAC,

maximal frequency = 2 kHz.

5

Ground for pulse output

6

Analog input 1, for temperature
measurement.

0(4)...20 mA, R

i

= 58,2 Ω, fuse = 50 mA.

7

Common ground

INFORMATION!
The current output of the UFC 030 F-EEx can be set according to NAMUR NE43. The current
output will go either to 3.6 or 21.5 mA in case of failure indication.