6.4.7 Plausibility filter (menu 3.07.09...3.07.11)

A plausibility filter can be set for the sensor measuring paths. There are three settings:

Error limit (menu 3.07.09)

Every measured value outside the error limit is not processed and will increase an internal
plausibility counter by 1, until a counter limit has been reached. The corresponding measuring
path will then be made inactive.

Range: 1 through 99%, default setting: 20%.

Error counter decrease step value (menu 3.07.10)

Every measured value within the error limit will decrease the internal plausibility counter by the
number programmed in this function. The higher the number, the faster an inactive measuring
path will become active again.

Range: 1 through 99, default setting: 4.

Error counter limit (menu 3.07.11

The counter limit is set using menu 3.07.11. Default setting is 0, i.e. the function is switched off.

INFORMATION!
Whenever one or more measuring paths are made inactive, the display compass needle will
indicate that, combined with the display of error messages.