Operation – KROHNE UFM 530 HT EN User Manual
Page 46
6
OPERATION
46
UFM 530 HT
www.krohne.com
12/2013 - 4002393201- HB UFM 530 HT R02 en
3.07.00, USER DATA
3.07.01
LANGUAGE
Language for display texts:
1) GB/USA: English.
2) D: German.
3) F: French.
3.07.02
ENTRY CODE
Entry code for setting mode:
1) NO: entry with → key only.
2) YES: entry with → key and code 1, factory set on 9 x key.
3.07.03
CODE 1
Press any 9-keystroke combination and then press the same
combination again. Each keystroke is acknowledged by "?" in the
display. If both combinations are equal, "CODE OK" appears and the
new code can be stored. Otherwise "WRONG CODE" appears and the
desired code has to be entered again.
3.07.04
LOCATION
Tag name setting. Free settable tag for identification, maximum
10 characters.
Characters assignable to each place: A...Z, blank character, 0...9.
Factory setting: KROHNE.
3.07.05
UNIT TEXT
Text for user-defined unit. Definition: volume/time.
Characters assignable to each place: A...Z, blank character, 0...9.
Note: fraction bar "/" in 5th place is unalterable!
Factory setting: XXXX/YYY.
3.07.06
UNIT VOL
User-defined unit volume. Quantity of user-defined volume per m
3
.
Range: 10
-5
...10
7
, factory setting: 1.
3.07.07
UNIT TIME
User-defined unit time. Amount of user-defined time in seconds.
Range: 10
-5
...10
7
, factory setting: 1.
3.07.08
RST ENABLE
Enable the reset of the display totalizer(s). Note: this function doesn't
affect the ability to reset the display totalizer(s) using the digital
input! Available options:
1) NO: totalizer reset disabled.
2) YES: totalizer reset enabled.
3.07.09
ERR LIMIT
Error limit in % of measured value for plausibility filter on the sensor
paths.
3.07.10
CNT DECR
Counter decrement setting for plausibility filter.
3.07.11
CNT LIMIT
Counter limit for plausibility filter. When "0" is set, the plausibility
filter will become inactive.
Range: 0...1000, factory setting: 0.