Operation – KROHNE UFM 530 HT EN User Manual

Page 46

background image

6

OPERATION

46

UFM 530 HT

www.krohne.com

12/2013 - 4002393201- HB UFM 530 HT R02 en

3.07.00, USER DATA

3.07.01

LANGUAGE

Language for display texts:

1) GB/USA: English.

2) D: German.

3) F: French.

3.07.02

ENTRY CODE

Entry code for setting mode:

1) NO: entry with → key only.

2) YES: entry with → key and code 1, factory set on 9 x key.

3.07.03

CODE 1

Press any 9-keystroke combination and then press the same

combination again. Each keystroke is acknowledged by "?" in the

display. If both combinations are equal, "CODE OK" appears and the

new code can be stored. Otherwise "WRONG CODE" appears and the

desired code has to be entered again.

3.07.04

LOCATION

Tag name setting. Free settable tag for identification, maximum

10 characters.

Characters assignable to each place: A...Z, blank character, 0...9.

Factory setting: KROHNE.

3.07.05

UNIT TEXT

Text for user-defined unit. Definition: volume/time.

Characters assignable to each place: A...Z, blank character, 0...9.

Note: fraction bar "/" in 5th place is unalterable!

Factory setting: XXXX/YYY.

3.07.06

UNIT VOL

User-defined unit volume. Quantity of user-defined volume per m

3

.

Range: 10

-5

...10

7

, factory setting: 1.

3.07.07

UNIT TIME

User-defined unit time. Amount of user-defined time in seconds.

Range: 10

-5

...10

7

, factory setting: 1.

3.07.08

RST ENABLE

Enable the reset of the display totalizer(s). Note: this function doesn't

affect the ability to reset the display totalizer(s) using the digital

input! Available options:

1) NO: totalizer reset disabled.

2) YES: totalizer reset enabled.

3.07.09

ERR LIMIT

Error limit in % of measured value for plausibility filter on the sensor

paths.

3.07.10

CNT DECR

Counter decrement setting for plausibility filter.

3.07.11

CNT LIMIT

Counter limit for plausibility filter. When "0" is set, the plausibility

filter will become inactive.

Range: 0...1000, factory setting: 0.