Operation – KROHNE UFM 530 HT EN User Manual

Page 50

background image

6

OPERATION

50

UFM 530 HT

www.krohne.com

12/2013 - 4002393201- HB UFM 530 HT R02 en

Error
message

Description of error message

What to do

ADC AN INP

Analog input internal error (A1or A2).

Switch off the flowmeter and switch it on

again. If the error still exists, contact your

local representative.

COMMUNIC

Communication device internal error.

Reset the error and wait for one minute. If

the error reappears, contact your local

representative.

CURR > MAX

Current output overflow (> 22 mA).

Check flow velocity.

DSP

Digital signal processor (DSP) internal

error.

Only checked at power-up. Switch off the

flowmeter and switch it on again. If the

error still exists, contact your local

representative.

EE MENU

Menu parameters corrupted.

Contact your local representative.

EE SERVICE

Service parameters internal error.

Contact your local representative.

EMPTY PIPE

Measuring tube not completely filled,

flowreading to 0, error on all paths.

Fill measuring tube completely.

FLOW > MAX

Measuring range overflow

(flow > 2 x Q

max

).

Check flow velocity.

FRONT END

Front end internal error.

Only checked at power-up. Switch off the

flowmeter and switch it on again. If the

error still exists, contact your local

representative.

INP1 < MIN

Analog input 1 too low (<3.6 mA).

Check analog input 1 connection.

INP1 > MAX

Analog input 2 too high (>22 mA).

Reduce analog input 1 current.

INP2 < MIN

Analog input 2 too low (<3.6 mA).

Check analog input 2 connection.

INP2 > MAX

Analog input 2 too high (>22 mA).

Reduce analog input 2 current.

RESTART

Flow meter restarted.

Reset errors.

UNRELIABLE

Flow data disturbed, same as triangle 4 in

the compass field.

Check flow conditions.

OPEN CIRC

Sensor X not connected or damaged

(combined with message SENSOR X).

Check connection of sensor X. For sensor

numbering see description of test function

2.03.04.

PATH 1

Measuring path 1 error.

Check flow conditions.

PATH 2

Measuring path 2 error.

Check flow conditions.

PULS > MAX

Pulse output overflow (>120%).

Check flow velocity.

SENSOR X

Sensor X error (combined with OPEN CIRC

or SHORT CIRC message).

Check connection sensor X.

SHORT CIRC

Sensor X short-cicuited (combined with

message SENSOR X).

Check connection sensor X.

TIME/DATE

Real time clock internal error.

Not available, reserved for future use.

TOT > DISP

Totalizer out of display range (maximum:

eight characters).

Reset totalizer or change totalizer unit.

TOT CHKSUM

Totalizer corrupted.

Reset totalizer.

UP2

μP2 internal error.

Contact your local representative.