Operation – KROHNE UFM 530 HT EN User Manual
Page 50
6
OPERATION
50
UFM 530 HT
www.krohne.com
12/2013 - 4002393201- HB UFM 530 HT R02 en
Error
message
Description of error message
What to do
ADC AN INP
Analog input internal error (A1or A2).
Switch off the flowmeter and switch it on
again. If the error still exists, contact your
local representative.
COMMUNIC
Communication device internal error.
Reset the error and wait for one minute. If
the error reappears, contact your local
representative.
CURR > MAX
Current output overflow (> 22 mA).
Check flow velocity.
DSP
Digital signal processor (DSP) internal
error.
Only checked at power-up. Switch off the
flowmeter and switch it on again. If the
error still exists, contact your local
representative.
EE MENU
Menu parameters corrupted.
Contact your local representative.
EE SERVICE
Service parameters internal error.
Contact your local representative.
EMPTY PIPE
Measuring tube not completely filled,
flowreading to 0, error on all paths.
Fill measuring tube completely.
FLOW > MAX
Measuring range overflow
(flow > 2 x Q
max
).
Check flow velocity.
FRONT END
Front end internal error.
Only checked at power-up. Switch off the
flowmeter and switch it on again. If the
error still exists, contact your local
representative.
INP1 < MIN
Analog input 1 too low (<3.6 mA).
Check analog input 1 connection.
INP1 > MAX
Analog input 2 too high (>22 mA).
Reduce analog input 1 current.
INP2 < MIN
Analog input 2 too low (<3.6 mA).
Check analog input 2 connection.
INP2 > MAX
Analog input 2 too high (>22 mA).
Reduce analog input 2 current.
RESTART
Flow meter restarted.
Reset errors.
UNRELIABLE
Flow data disturbed, same as triangle 4 in
the compass field.
Check flow conditions.
OPEN CIRC
Sensor X not connected or damaged
(combined with message SENSOR X).
Check connection of sensor X. For sensor
numbering see description of test function
2.03.04.
PATH 1
Measuring path 1 error.
Check flow conditions.
PATH 2
Measuring path 2 error.
Check flow conditions.
PULS > MAX
Pulse output overflow (>120%).
Check flow velocity.
SENSOR X
Sensor X error (combined with OPEN CIRC
or SHORT CIRC message).
Check connection sensor X.
SHORT CIRC
Sensor X short-cicuited (combined with
message SENSOR X).
Check connection sensor X.
TIME/DATE
Real time clock internal error.
Not available, reserved for future use.
TOT > DISP
Totalizer out of display range (maximum:
eight characters).
Reset totalizer or change totalizer unit.
TOT CHKSUM
Totalizer corrupted.
Reset totalizer.
UP2
μP2 internal error.
Contact your local representative.