6.4 Important menu functions in detail
In this chapter the different functions of the menu structure are described in more depth.
6.4.1 Error/totalizer (Menu 0.00.00)
This menu is accessible from the measuring mode by pressing the ^ key to enter "CODE 2" and
afterwards pressing the key ↑ and and then →.
Depending on the programming of function 3.03.07, the error messages alternate with the
display of the measured value(s) every five seconds or they can be manually selected by pressing
the ↑ key.
The following list gives an alphabetical overview of error messages that can occur during
process flow measurement and what to do:
INFORMATION!
Since the UFC 030 converter can be equipped with different options, the availability of certain
options depends on the function of the converter.
INFORMATION!
Only when function 3.03.08 is set to YES, errors occurring during process flow measurement are
represented with flashing display lines and/or a compass field!
1 Compass needle, indicating measuring path error(s): depending on the version of the sensor triangle a appears when
there is no measured value from path 2 and triangle b from path 1; triangle c appears when there is too much noise
on the measuring path(s) for proper measurement.
2 Flashing bar, indicating "new" errors, i.e. not yet acknowledged.
3 Flashing line with description of error message(s).
4 Flashing line with number of errors that have occurred.