LIST OF
WARNINGS AND CAUTIONS
BOP 1KW OPR 8/7/09
G/(H Blank)
PAGE
WARNING/CAUTION
When the ADJUST control is rotated, the active parameter is immediately
effective if the output is enabled (on = STANDBY indicator not lit). The
voltage/current applied to the load changes as the ADJUST control is ro-
When working with active loads, always adjust the BOP protection limits
to be above the maximum values of voltage or current expected from the
load. For example, when the BOP is operating in voltage mode sinking
energy from a constant current type load, set the current protection limits
of the BOP above the maximum current expected from the load.
to enter the Saved Waveform Screen from the power-
up screen, perform the following steps to avoid unexpected voltages or
currents from previous settings being briefly applied to the load when the
waveform is first executed.
Pressing RESET will cause voltage transients to appear at the output
which may damage a connected load. Switch is recessed to prevent inad-
When the serial port has received an XOFF, the error message -400, QUE
error will be placed in the queue to indicate the loss of transmitted infor-
mation due to a received XOFF character. When XON is received, the unit
will transmit all data in it's buffer followed by the exclamation character
(!). This (!) character is not part of any message from the BOP and indi-
cates the transmission buffer has been cleared and the BOP is idle.
function is available, perform this function
first, as it affects all other output calibrations.