3-44

BOP HIPWR 080709

3.5.3.1

GPIB PORT SETUP

The following paragraphs describe the how to configure the GPIB port.

3.5.3.1.1 CHANGING THE GPIB ADDRESS

The default address is 6. To change the GPIB address press

%

from the power-up screen,

then press

!

to enter the Interface menu (see Table 3-16) Highlight GPIB Address and press

!

, then use the number keys to enter the desired address and press

$

to save. Press

$

to

save for power-up or

%

to apply the changes (without saving for power-up) and exit.

3.5.3.1.2 CONFIGURE DEVICE CLEAR (DCL) CONTROL

The device clear (DCL) and selected device clear can be set to operate in two modes. In the
MATE mode, when the device clear is received, the output of the power supply is set to zero
volts. In the SCPI mode sending DCL or selected DCL has no effect on output voltage and cur-
rent as required by IEEE specification 488.2. The factory default value is SCPI mode.

To change the Device Clear configuration press

%

from the power-up screen, then press

!

to

enter the Interface menu (see Table 3-16) Highlight Device Clear and press

!

, then highlight

SCPI or MATE and press

$

to save. Press

$

to save for power-up or

%

to apply the

changes (without saving for power-up) and exit.

TABLE 3-16. INTERFACE SETTINGS MENU

SETTING

CHOICES

(BOLD = Factory Default)

FUNCTION

Interface
Data Format

SCPI

CIIL

SCPI (Standard Commands for Programmable Instruments) - see PAR.
3.6. and Appendix B.
CIIL - Provides compatibility with early Kepco power supplies and con-
trollers.

GPIB Address

0 - 30 (06)

Assigns address used by external GPIB controller to communicate with
BOP.

*RST set Output

OFF

ON

OFF - Sending *RST causes output to go off (per IEEE 488.2 specifica-
tions) (see PAR. 3.5.3.1.3 for details).
ON - Sending *RST causes output to go on (see PAR. 3.5.3.1.3 for
details).

Device Clear

SCPI

MATE

SCPI - sending Device Clear (DCL) or selected DCL has no effect on
output voltage and current per IEEE specification 488.2.
MATE - When device clear is received, the power supply output is set to
zero volts.

Serial Baud Rate

OFF

9600

19200

Selects RS 232 serial port baud rate.

Xon/xoff

ENABLE

DISABLE

ENABLE - Allows BOP to control when data is received (see PAR.
3.5.6.2.1).
DISABLE - Disables XON/XOFF. If both Serial XON/XOFF and Prompt
functions are disabled, Echo mode is the default (see PAR. 3.5.6.2.2).

Prompt

DISABLE

ENABLE

ENAB+ECHO

ENABLE - Enables Serial Prompt (see PAR. 3.5.6.2.3).
DISABLE - Disables Prompt. If both Serial XON/XOFF and Prompt
functions are disabled, Echo mode is set (see PAR. 3.5.6.2.2).
ENAB+ECHO - Enables Serial Prompt (see PAR. 3.5.6.2.3) and Echo
(see PAR. 3.5.6.2.2) modes. Prompt is returned when the unit is ready
and any received characters are echoed back to the sender.

NOTE: To restore all interface settings defaults, press

%

from the power-up screen, highlight Interface Settings, press

!

, then

press

@

to restore defaults shown above in BOLD. Press

$

to save for power-up or

%

to apply the changes

(without saving for power-up) and exit.