3 preliminary operational check, Caution, A-c wiring (e.g., defeating of ground connection) – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.2.38 to 2.47 User Manual

Page 44: Preliminary operational check -6, 10 analog i/o port input/output pin assignments -6, Able 2-10), Able 2-10, R. 2.3, Able 2-10.)

background image

2-6

BOP HIPWR 080709

2.3

PRELIMINARY OPERATIONAL CHECK

A simple operational check after unpacking and before equipment installation is advisable to
ascertain whether the power supply has suffered damage resulting from shipping.

Refer to Figures 2-1 and 3-1 for location of operating controls and electrical connections. Tables
3-1 and 3-2 explain the functions of operating controls/indicators and keypad keys, respectively.
Refer to PAR. 3.2 for a description of basic operating techniques.

TABLE 2-10. ANALOG I/O PORT INPUT/OUTPUT PIN ASSIGNMENTS

CONNECTOR

PIN

SIGNAL NAME

FUNCTION

ANALOG I/O

PORT

A2A5J6

CAUTION: IT IS RECOMMENDED THAT SOURCE POWER OF EXTERNAL EQUIPMENT CON-

NECTED TO THE ANALOG PORT BE APPLIED THROUGH AN ISOLATING TRANS-
FORMER TO AVOID GROUND LOOPS OR POSSIBLE DAMAGE TO THE BOP DUE TO
INCORRECT EQUIPMENT A-C WIRING (E.G., DEFEATING OF GROUND CONNECTION).

1

NC

2

VM-/CM

External input signal, TTL logic referenced to pin 9, controls the mode
of operation when using external reference (goes to the digital board
that changes VM-/CM signal). Logic 0 (or pin grounded) = current
mode, logic 1 (or pin not connected, the default) = voltage mode (see
PAR. 3.4.2).

3

IOUT_DMM

Output analog signal through 2K ohm resistor referenced to pin 4 for
monitoring overall output current, 0V to ± 10V corresponds to zero to ±
full scale current.

4

SGND

Signal Ground used for IOUT_DMM (pin 3) return.

5

– I_LIM_EXT

Analog input signal referenced to pin 12, 0V to +10V sets the negative
current limit between zero and –I

Omax

, +10V corresponds to rated

maximum current (e.g., for BOP 36-28MG +10V sets negative current
limit to –28A) (see PAR. 3.4.4). Open = disabled (see NOTE 1).

6

– V_LIM_EXT

Analog input signal, 0V to +10V, sets the negative voltage limit
between zero and –E

Omax

, +10V corresponds to rated maximum volt-

age (e.g., for BOP 36-28MG +10V sets negative voltage limit to –36V).
Open = disabled. (See PAR. 3.4.4.) Open = disabled (see NOTE 1).

7

NC

8

NC

9

GND

Ground - Used for VM-/CM (pin 2) return

10

SGND

Signal Ground (used for EXT_REF (pin 11) return

11

EXT_REF

External analog reference signal referenced to pin 10, used for both
voltage mode and current mode, 0V to ±10V corresponds to zero to ±
rated nominal (full scale), voltage or current (see PAR. 3.4.3).

12

GND1

Ground (Used for pin 5, 6, 13 and 14 return)

13

+I_LIM_EXT

Analog input signal referenced to pin 12, 0V to +10V sets the positive
current limit between zero and I

Omax

; +10V corresponds to rated maxi-

mum current (e.g., for BOP 36-28MG +10V sets positive current limit to
+28A) (see PAR. 3.4.4). Open = disabled (see NOTE 1).

14

+V_LIM_EXT

Analog input signal, 0V to +10V sets the positive voltage limit between
zero and E

Omax

, +10V corresponds to rated maximum voltage (e.g.,

for BOP 36-28MG +10V sets positive current limit to +36V. Open = dis-
abled. (See PAR. 3.4.4.) (see NOTE 1).

15

NC

NOTE 1. When disabled, the external limit channels are automatically set 20% higher than BOP nominal references.