BHK-1/2-MG 121313

Failure to provide an adequate delay can result in:

• Commands that are not processed,

• The following command may be received in error, causing an error in the transmission,

• Unit lock-up requiring power cycling of the unit. If working via the GPIB bus, sending

Interface Clear and Device Clear followed by *RST will unlock the unit.

FIGURE 3-2. PROGRAMMING EXAMPLE TO VERIFY PREVIOUS COMMAND HAS COMPLETED

3.4

BHK-MG VISA INSTRUMENT DRIVER

The VISA instrument driver simplifies programming with a VISA compatible GPIB controller. and
Includes 1) source code (C) for all VISA functions, and 2) a complete programming reference

#include

#include

#include

#include

/*Overhead for the use of a NATIONAL INSTRUMENTS gpib interface */

int unit_desc;

// handle for the national instruments controller

int GPIbus=0;

// GPIB card 0

int adr=6;

// Power Supply address

char status_byte;

// status byte from the power supply

#define MAV 0x10 /* bit 4 of the status byte is the Message AVailable bit by 488.2 specification */

/* Function Send_with_wait

INPUT: string to be sent to power supply

Description: adds the *OPC? query and performs serial polls to wait for the command to be completed.

*/

int Send_with_wait(char *command);

char snd[501];

// data to be sent to the power supply

char rcv [10];

// data from power supply

int j;

sprintf(snd,”%s;:*OPC?,command);

// Add *OPC? to the command

// so there is a response from the

// power supply

Send(GPIbus, adr, snd, strlen(snd), 2);

// Send the data to the power supply

for (j=0;j<500;j++)(

// loop until ready (5 seconds max)

Delay(.05);

// Wait for command to complete

ibrsp(unit_desc,&status_byte);

// get status byte

if ((status_byte& 0x10) ==0x10) break;)

// by looking for data in string

Receive (GPIbus, adr, rev, rev_buf_size,10);

// so the error queue will not receive a 410 error

}

main(

// test code to show operation of function.

unit_desc=ibdev(GPIbus,adr,adr/256,T100ms,1,0x40a);Delay(.005);

Send (GPIbus,adr,”VOLT 10;curr .01”,sizeof(“VOLT 10;curr .01”),NLEND;

Send_with_wait(“*SAV 10”);

}