B.20 [SOURce:]LIST:COUNt:SKIP? QUERY

LIST:COUN:SKIP?

Syntax:

Short Form: LIST:COUN:SKIP? Long

Form:

LIST:COUNt:SKIP?

Return Value:

Description: Identifies how many steps will skipped the first time the list is executed. Returns value set by

LIST:COUN:SKIP command. (See example, Figure B-3.)

B.21 [SOURce:]LIST:CURRent COMMAND

LIST:CURR

Syntax:

Short Form: LIST:CURR , , . . . (to max of 250 data points)
Long Form: LIST:CURRent , , . . . (to max of 250 data points)
= digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Adds the current value (in Amps) to list. This command sequentially adds LIST:CURRent values to

the main channel List Data Table. Starting location is indicated by LIST:CURR:POIN? The maximum
number of entries is 250. Since the input buffer of the BHK has a limit of 253 characters, multiple com-
mands are necessary to complete the full 250 entries of the list. If LIST:CURR is entered only for loca-
tion 0, that current will apply to all steps when VOLT:MODE LIST is executed. Related Commands:
LIST:CURR:POIN?. (See example, Figure B-3.)

B.22 [SOURce:]LIST:CURRent? QUERY

LIST:CURR?

Syntax:

Short Form: LIST:CURR?

Long Form: LIST:CURRent?

Return Value: , , . . . to

Description: Identifies the parameters (main channel) entered for the list. Starting at location established by

LIST:QUERy, returns comma-separated list of up to 16 values indicating the main channel parameters
entered. i.e., the contents of main channel List Data Table. Related Commands: LIST: CURR,
LIST:QUERy. (See example, Figure B-3.)

B.23 [SOURce:]LIST:CURRent:POINts? QUERY

LIST:CURR:POIN?

Syntax:

Short Form: LIST:CURR:POIN? Long

Form:

LIST:CURRent:POINts?

Return Value: (0 to 250)

Description: Identifies the total number of points in a list and the next location to be filled by LIST:CURR

command. The LIST:CURR pointer is initially at 0 via LIST:CLE. For each data point entered by a
LIST:CURR command the list pointer is incremented If LIST:CURR:POIN? returns 5, the LIST:CURR
pointer is at 5 indicating there are 6 data points comprising the list. Related Commands: LIST:CURR.
(See example, Figure B-3.)

B.24 [SOURce:]LIST:DIRection COMMAND

LIST:DIR

Syntax:

Short Form: LIST:DIR (UP|DOWN)

Long Form: LIST:DIRection (UP|DOWN)

Description: Allows the list to be executed from beginning to end (UP) or from end to the beginning

(DOWN). LIST:CLEar sets the list to the UP direction (beginning to end). Related Commands:
LIST:DWEL?. (See LIST example, Figure B-3.)

B.25 [SOURce:]LIST:DIRection? QUERY

LIST:DIR?

Syntax:

Short Form: LIST:DIR?

Long Form: LIST:DIRection?

Return Value: (1 or 0)

Description: Identifies the direction (1 = up or 0 = down) for executing the list established by LIST:DIR.

Related Commands: LIST: DIR. (See example, Figure B-3.)

B.26 [SOURce:]LIST:DWELl COMMAND

LIST:DWEL

Syntax:

Short Form: LIST:DWEL (0.010 to 655.35),,, . . . to maximum of 250 values
Long Form: LIST:DWELl (0.010 to 655.35),,, . . . to maximum of 250 values

Description: Determines how long the main channel parameters will be active. Sets time value (from 0.010 to

655.35) in seconds for List:Dwell locations of the List Data Table. The main channel (either voltage or
current) is determined by the load. If LIST:DWEL is entered for only location 0, that time duration will
apply to all steps when either VOLT:MODE LIST or CURR:MODE LIST is executed. Related Com-
mands: VOLT:MODE, LIST:CURR, LIST:VOLT, LIST:DWEL? (See example, Figure B-3.)