2.12

TRACe

Subsystem

2-25

2.13 SYSTem subsystem

GW APS-1102 SYST Beep.vi
This controls on/off of beep sound.

Beep

This sets beep sound on/off. The default setting is “ON”.

Query Beep

When Read = True, outputs the beep sound status of the APS-1102. When Read = False,
“space” is set.

GW APS-1102 SYST Extio.vi
This controls the external control input enable/disable setting.

Extio

This sets external control input on/off. The default setting is “ON”.

Query Extio

When Read = True, outputs the external control input status of the APS-1102. When Read =
False, “space” is set.

GW APS-1102 SYST Err.vi
This reads one error from the error queue (FIFO), and clears the read error in the error queue. The

maximum number of errors that can be held in the error queue is 20. If 21 or more errors occur, [-350,
“Too Many Errors”] is inserted as the 20th error.

Error Number

This reads and clears an error from the error queue (FIFO).

GW APS-1102 SYST Pon.vi
This sets the output status at power-on. When the output status is on at power-on, the output turns

on after the start-up and its display finishes.

However, if the system is restarted after a problem has been detected and the system has been locked,

it will be started with output off regardless of this setting.