APS-1102 LabVIEW Driver Manual

2-14

GW APS-1102 MEAS Curr Res.vi
This resets the stored output current peak value (hold) to 0 Apk.

GW APS-1102 MEAS Curr Ave.vi
This returns the output current average value (AC+DC average value) in A units.

Average

This outputs the output current average value (AC+DC average value).

Error

When a measurement error occurs, “True” is set.

GW APS-1102 MEAS Curr Cres.vi
This returns the output current crest factor.
The crest factor is calculated as “current peak value/current RMS value”.

Average

This outputs the crest factor (peak-to-rms ratio) of output current.

Error

When a measurement error occurs, “True” is set.

GW APS-1102 MEAS Curr Harm.vi
This returns the harmonic current [Arms] in the specified degree range.
However, when in ACT-INT mode, if the frequency neither 50 Hz nor 60 Hz, the harmonic current

cannot be measured. ([3, “Invalid in this mode”] or [6, “Invalid in this frequency”] error occurs.)

Parameter

This specifies the order range to be outputted. Any of 1 (1st to 10th)/2 (11th to 20th)/3 (21st
to 30th)/4 (31st to 40th) can be specified. The default setting is “1”.

Output [Arms]

This outputs the harmonic current [Arms] in the specified order range in an array. When the
order range specified in Parameter is “1”, 1st, 2nd, … 10th are output from the top.

Error