Instrument shut-down, Daily shutdown, Complete shutdown – Agilent Technologies G6600-90006 User Manual
Page 77

Operation and Maintenance Manual
77
Instrument Shut-Down
Daily Shutdown
1
Toggle off the ozone generator.
2
Turn off the air regulator (counter-clockwise), located inside Detector door.
3
Toggle power to "stand-by."
4
Leave the vacuum pump and Dual Plasma Controller operating at all times.
Complete Shutdown
1
Toggle off the ozone generator.
2
Turn off the air regulator (counter-clockwise), located inside Detector door.
3
Toggle power to STANDBY.
4
Turn off power to the Dual Plasma Controller.
5
After 15 minutes, toggle off the vacuum pump, so the Burner cools and
moisture is removed from the system.
6
Turn off gases to unit.