Instrument Shut-Down

Daily Shutdown

1

Toggle off the ozone generator.

2

Turn off the air regulator (counter-clockwise), located inside Detector door.

3

Toggle power to "stand-by."

4

Leave the vacuum pump and Dual Plasma Controller operating at all times.

Complete Shutdown

1

Toggle off the ozone generator.

2

Turn off the air regulator (counter-clockwise), located inside Detector door.

3

Toggle power to STANDBY.

4

Turn off power to the Dual Plasma Controller.

5

After 15 minutes, toggle off the vacuum pump, so the Burner cools and
moisture is removed from the system.

6

Turn off gases to unit.