Plugged restrictor lines – Agilent Technologies G6600-90006 User Manual

Page 110

background image

110

Operation and Maintenance Manual

Plugged Restrictor Lines

Plugged restrictor lines are verified by turning off the air/oxygen regulator
located inside the door of the Detector and observing little or no change in the
pressure as indicated on the regulator. The plug can be located in either the
pre-ozone restrictor or the post-ozone restrictor (see

Figure 3

). A partially

plugged restrictor line can result in a non-linear signal, or reduced sensitivity
which will vary with the flow of ozone. Replace the restrictor lines.