Operation and Maintenance Manual
Plugged Restrictor Lines
Plugged restrictor lines are verified by turning off the air/oxygen regulator
located inside the door of the Detector and observing little or no change in the
pressure as indicated on the regulator. The plug can be located in either the
pre-ozone restrictor or the post-ozone restrictor (see
). A partially
plugged restrictor line can result in a non-linear signal, or reduced sensitivity
which will vary with the flow of ozone. Replace the restrictor lines.