beautypg.com

Detector operation, Detector stability and response, Column bleed – Agilent Technologies G6600-90006 User Manual

Page 73: Coking, Hydrogen poisoning

background image

Operation and Maintenance Manual

73

Detector Operation

Detector Stability and Response

The time required for system stabilization varies depending on the application,
system cleanliness, presence of active sites and other factors. Useful results
could be generated within 30 minutes of start-up, especially with a previously
operated system. A longer stabilization time is likely to be required upon
changing critical system components, such as the combustion tubes or the GC
column. In addition, gas flow rates may drift initially as thermal equilibrium is
reached due to changes in gas viscosity with temperature. Therefore, it is good
practice to monitor gas flow rates and adjust them accordingly.

Even though a system may not be fully stabilized, sample injections can be
made within minutes of instrument start-up. Whether the results are useful
largely depends on application. Typically, an elevated baseline will initially be
observed, which will diminish upon successive programmed runs.

After stabilization has been reached, the system should exhibit good
short-term and long-term precision. Of course this also depends on the
application and concentration of components being measured. As an example,
analysis of thiophene in benzene at the 1 ppm sulfur level yielded 1.4% RSD
(n=10) over about 2 hours and 3.6% RSD (n=42) over about 96 hours. As
expected, carbon disulfide at a lower concentration of 90 ppb sulfur yielded
2.6% and 10.4% RSD, respectively.

Column Bleed

Accumulation of column bleed causes silicon dioxide to build up in the Burner.
This silicon dioxide creates active sites that are detrimental to performance. In
many cases, the choice of column can be optimized for a particular
application. Column bleed can be minimized by the use of oxygen traps on the
carrier gas, low-bleed columns, and lowest possible maximum oven
temperature.

Coking

Contamination from some sample matrices can reduce sensitivity. Crude oils
containing volatile metal complexes may contaminate ceramic tubes. The
incomplete combustion of certain hydrocarbon-containing compounds leaves
behind coke deposits on the tubes. Coke deposits may be removed from the
Burner by reducing the hydrogen flow rate. The Dual Plasma Burner is much
less susceptible to coke formation than other designs.

Hydrogen Poisoning

Hydrogen poisoning of the ceramic tubes occurs when there is no oxidizer flow
through the ceramic tubes. The result is extremely reduced, or no response.
Hydrogen poisoned tubes can not be reconditioned and should be discarded.