Table 7. troubleshooting burner issues – Agilent Technologies G6600-90006 User Manual

Page 116

background image

116

Operation and Maintenance Manual

Table 7 Troubleshooting Burner Issues

Problem

Possible Cause

Diagnosis

Corrective Action

Fuses Blow on Startup

Shorted heater element.

Inspect for exposed wires or
broken insulation around
wire leads. Measure
resistance leads, should not
be 0.

Take Burner top off and
reseat insulation around
heater leads.

Low Sensitivity

Leaks at ferrules or fittings.

For SCD, use a microliter
syringe containing a small
amount of sulfur compound
e.g., CS

2

to “snoop” for leaks

while watching mV output.
Leaks are very evident by a
large increase in signal
displayed on the LED.

For NCD, a high background
could be observed from
atmospheric nitrogen.

Inspect fittings for scoring
and replace if necessary.
Replace ferrules if worn.
Hand tighten all fittings
mounted onto vespel
ferrules, and wrench tighten
all connections with Valco
ferrules.

Low Sensitivity

Improper gas flows.

Measure gas flows.

Reset gas flows.

Low Sensitivity

High Pressure

Broken ceramic tubes or
ferrules

Remove and inspect tubes
and ferrules for chips and/or
breaks.

Replace tubes or ferrules.

Low Sensitivity

High Background

Improper column
positioning.

Inspect for charred Detector
end of column.

After cutting off charred end
of column, measure column
length 108-109 mm from
ferrule (or 114-115 mm from
bottom of nut) and re-attach
column.