Dual Plasma Controller Controls

The Agilent Dual Plasma Controller provides easy access to basic settings.

The baseline signal from the SCD and NCD can be
adjusted from 0 to ±1% of the full scale recorder
output using the offset control. (e.g., +10 mV to
-10 mV for a 1 V full scale setting). This control
can be used to offset the background signal.

Figure 22 Dual Plasma Controller Front Panel

Oxidizer and
Hydrogen Control
Knobs

Allow you to adjust the oxidizer and hydrogen flow
rate.

Temperature Control
Knob

Allows you to adjust the Burner temperature.

Selector Control Knob

Changes the display to show the current value or
set point for each setting (temperature in °C,
pressure in Torr, oxidizer in sccm, or hydrogen in
sccm).