Operation and Maintenance Manual

3

Remove the old fuses from the holder and replace with new fuses.

If the Controller still does not power on after replacing the fuse, contact
Agilent.

Consult

Table 4

for the appropriate replacement fuse.

Table 4 Fuses for 100 V, 120 V and 230 V Versions of 355 SCD and 255 NCD

100/120 V

230 V

Detector Back Panel

Main AC Power

15 A/250 V

3AG slo-blo

T10 A/250 V

5 x 20mm

Detector

Electronics Power Supply (F1)

250 mA/250 V

3AG slo-blo

T125 mA/250 V

5 x 20mm

Pump (F2)

15 A/250 V

3AG slo-blo

T5 A/250 V

5 x 20mm

Ozone Generator (F3)

1 A/250 V

3AG slo-blo

T100 mA/250 V

5 x 20mm

Photomultiplier Tube Cooler

1 A/250V
3AG slo-blo

T500 mA/250V
5 x 20mm

Dual Plasma Controller Back Panel

Main AC Power (2 fuses)

T2 A/250 V

5 x 20mm

1 A/250 V

5 x 20mm

Figure 47 Fuse Positions on the Power Supply Board

OZONE

PUMP

F1

F3

F2

ELECTRONICS

Side View

ELECTRONICS

OZONE

PUMP

T125 mA

Front View (230 V Unit)

T100 mA

T5 A