beautypg.com

Specifications, Technical information - 355 scd, Technical information - 255 ncd – Agilent Technologies G6600-90006 User Manual

Page 20: Technical information — 355 scd, Technical information — 255 ncd

background image

20

Operation and Maintenance Manual

Specifications

Technical Information — 355 SCD

Technical Information — 255 NCD

Sensitivity

*

* Burner temperature 800 °C, 80 SCCM air, and 60 SCCM hyrdrogen, test compound dimethyl sulfide in toluene.

Typical < 0.5 pg S/second (signal to noise 3.3:1)

Typical Selectivity

g S/g C > 2 x 10

7

Linearity

>10

4

Precision and Stability

† Based on thiopene in benzene at 1 ppm mass sulfur, 1 µL injection split 1:10, 30 m, 0.32 mm ID, 1 µm thick CP

wax (n=10 for 2.5 hours; n=42 for 72 hours).


Subject to change without notice.

<2% RSD 2 hours

<5% RSD 72 hours

Ozone flow through the Post Ozone Restrictor

20-30 mL/min at 3-6 psig

Typical reaction cell pressure

4 - 8 Torr RV5 Oil Sealed Pump

6 - 12 Torr Dry Piston Pump

Typical Burner Pressure

250-400 Torr operating

Typical Burner Temperature

800 °C

Typical Air Flow Rate

65 SCCM recommended

3-10 SCCM recommended with FID adapter

Typical Hydrogen Flow Rate

40 SCCM recommended

Signal Output Ranges

0-1V, 0-10V

Typical time to reach 800 °C from room
temperature

10 min typical(120 VAC, 60 Hz)

Typical safety shroud outside temperature

<65 °C at 800 °C Burner temperature typical

Recorder output

0-1 V or 0-10 V

Sensitivity

<3 pg N/second (signal to noise 3:1) in both N
and nitrosamine modes

Selectivity

g N/g C > 2 to 10

7

in N mode (selectivity in

nitrosamine mode is matrix-dependent)

Linearity

>10

4