Detection limits – Agilent Technologies G6600-90006 User Manual

Page 76

background image

76

Operation and Maintenance Manual

Detection Limits

The following table lists the detection limits which can be expected for typical
chromatographic conditions, assuming proper operation of the Detector and
chromatographic systems.

Table 2 Expected Detection Limits for Chromatographic Conditions

Type of Injection

Volume

Column

Detection Limit Per
Compound as Sulfur

Liquid, Split 1:10

1 µL

Capillary

≤0.32 mm ID 50 parts per billion

Liquid, Split 1:100

1 µL

Capillary

≤0.32 mm ID 0.5 parts per million

Liquid, On-column or splitless

1 µL

Capillary

≤0.32 mm ID 5 parts per billion

Liquid, Splitless

10 µL

Capillary

≤0.32 mm ID 0.5 parts per billion

Gas, Split 1:10

1 cm

3

Capillary

≤0.32 mm ID 50 parts per billion

Gas, Direct on-column

1 cm

3

0.53 mm ID

5 parts per billion

Gas, Direct on-column

1 cm

3

Packed

5 parts per billion

Gas, Direct on-column

10 cm

3

Packed

0.5 parts per billion