Detection limits – Agilent Technologies G6600-90006 User Manual
Page 76
76
Operation and Maintenance Manual
Detection Limits
The following table lists the detection limits which can be expected for typical
chromatographic conditions, assuming proper operation of the Detector and
chromatographic systems.
Table 2 Expected Detection Limits for Chromatographic Conditions
Type of Injection
Volume
Column
Detection Limit Per
Compound as Sulfur
Liquid, Split 1:10
1 µL
Capillary
≤0.32 mm ID 50 parts per billion
Liquid, Split 1:100
1 µL
Capillary
≤0.32 mm ID 0.5 parts per million
Liquid, On-column or splitless
1 µL
Capillary
≤0.32 mm ID 5 parts per billion
Liquid, Splitless
10 µL
Capillary
≤0.32 mm ID 0.5 parts per billion
Gas, Split 1:10
1 cm
3
Capillary
≤0.32 mm ID 50 parts per billion
Gas, Direct on-column
1 cm
3
0.53 mm ID
5 parts per billion
Gas, Direct on-column
1 cm
3
Packed
5 parts per billion
Gas, Direct on-column
10 cm
3
Packed
0.5 parts per billion